Digunakan untuk melakukan pengujian kualitas air
| Measurement range of the Professional COD Analyzer | : COD: 5 ~ 2000mg / L, exceeding the dilution determined |
| Indication error | : COD: Indication error ≤ ± 5% |
| Repeatability | : ≤ 3% |
| Resistance to chloride interference | : ≤ 2000mg / L |
| Temperature control system | : room temperature ~ 180 ℃ can be set, COD digestion temperature of 165 ℃ |
| Temperature control accuracy | : ± 1 ℃ |
| Temperature control time | : 1 ~ 999adjustable |
| Digestion time | : 10min |
| Optical stability | : Instrument absorbance values drift less than 0.002A within 20min |
| Batch capacity | : four water samples (or optional 16 water samples) |
| Dimensions | : COD analyzer 80mm × 230mm × 55mm |
| Digestion device | : 105mm × 160mm × 90mm |
| Weight | : 0.5kg digestion device |
| Host power consumption | : current <40μA; digestion power consumption: <65W |
| Ambient temperature | : 5 ~ 40 |
| Relative humidity | : ≤ 85% |
| Power supply | : Host: 4 piece 5 # batteries |
| Digestion device | : Car power adapter or 220V power supply(battery or optional power supply) |
| Gross Weight | : 13 KG |